A3076, Active Oscilloscope Probe, 10:1, 750 MHz
Bandwidth | DC to 750MHz into 50 Ohm load DC to 500MHz (nominal) system bandwidth with PicoScope 6000E 500MHz oscilloscope |
Rise time | < 466ps |
DC gain accuracy | ±3% of full scale over +15 to +40°C, after 20-minute warmup |
DC offset accuracy |
±3mV over +15 to +40°C, after 20-minute warmup |
Input characteristics | 1MOhm || 0.9pF |
Attenuation | 10:1 |
Dynamic performance (typical) | |
Noise | 2.5mV (RMS) referred to probe input |
Pulse response | < 10% overshoot |
Bandwidth flatness | (+1dB, –3dB) from DC to full bandwidth |
Low-frequency flatness | < ±3% (or ±0.3dB) from DC to 250MHz |
Propagation delay | 5.4ns |
Input voltage | |
Full scale | ±5V (DC + AC peak) |
Absolute max. voltage | ±30V (DC + AC peak) subject to frequency derating |
DC offset range | ±10V |
DC offset control accuracy | ±1% of offset setting, additional to DC accuracy above |
General | |
Probe detection | Automatic with any oscilloscope with Pico Intelligent Probe Interface (IPI) |
Operating temperature range | 0 to +50°C For quoted accuracy:15 to 40°C after 20-minute warmup. |
Storage temperature range | -20 to +60°C. |
Operating humidity range | 5% to 80% RH non-condensing |
Storage humidity range | 5% to 95% RH non-condensing |
Altitude range | Up to 2000m |
Pollution degree | EN 61010 degree 2 : “only nonconductive pollution occurs except that occasionally a temporary conductivity caused by condensation is expected” |
Dimensions (probe body) | 108mm (L) x 16mm (H) x 12mm (W) |
Cable length | 1.2m |
Safety compliance | EN 61010-031:2015 |
EMC compliance | Tested to EN61326-1:2013 and FCC Part 15 Subpart B |
Environmental approvals | RoHS, WEEE, REACH |